MY LABSHOP

PCR Cycler TurboCycler Lite

PCR Cycler TurboCycler Lite
Die Abbildung kann abweichen und Zubehör enthalten, das sich nicht im Lieferumfang dieses Artikels befindet.
  • (Klicken für mehr Informationen)
  • pdf_preview
  • pdf_preview
Lieferzeit auf Anfrage
Bestell-Nr.: TCLT9620 Preis: Login Haben Sie Fragen ? Rufen Sie 0043 (01) 244 28 55 oder senden Sie ein mail !
Specifications
Sample Block: Fixed 96-well, compatible with regular profile or low profile 0.2 ml PCR tube, strip, non-skirted, semi-skirted and full-skirted 96-well plate
Sample Volume: 10 - 100 µl
Block Temperature Range: 4.0 - 99.9 °C
Max. Heating Rate: 3.0 °C/sec
Max. Cooling Rate: 2.0 °C/sec
Temperature Accuracy: +/- 0.3 °C
Temperature Uniformity Across Block: +/- 0.5 °C
Adjustable Ramp Rate: High / Low
Slow Ramp Temperature Control: Via temperature increment / decrement between cycles

Gradient Direction: Horizontal across the block
Gradient Temperature Range: 30 - 99 °C
Gradient Temperature Difference Max. span: 24 °C
Heater Lid Temperature Setting Range: 35 - 120 °C or off
Temperature Accuracy: +/- 1.0 °C

Software
Stored Program No.: 200 sets
Registered User Folder No.: 50 sets
User Folder Password Protection: Yes

Display: Monochrome LCM, 240 x 128 pixels
Input Interface: Capacitive touch keypad
Auto Restart after Power Outage: Yes
Footprint Dimensions (H x W x D): 225 mm x 245 mm x 415 mm
Weight: 9.5 kg
Power Supply: AC 100-240 V, 50/60 Hz, 500 W
Certification: CE, RoHS

Features and Benefits
Intuitive Operation Experience - A sensitive capacitive touch keypad and an intuitive graphical interface
Gradient Optimization - The thermal gradient function allows fast PCR optimization for new experiments
Advanced Slow-Ramp Temperature Control - The ramp rate can be precisely controlled down to 0.1 °C/sec to meet the need for the CRISPR/Cas related assays
Fully Adjustable Lid Temperature - The temperature can be set between 35 and 120 °C for virtually any type of experiment including NGS pre-treatment
Easy Disinfection - The dust and aerosol proof keypad can be easily disinfected
Auto Restart - Power failure recovery keeps the experiment safe
mehr